Stage navigation function
(QS CNC model)
The stage navigation function allows X,Y stage movement with the mouse.
It speeds measurement and shortens part programming time by reducing
stage overrun and short run. For long travel, the user can click on the
drawing image in the graphic window; short travel is accomplished by clicking
on the video window.
Step feeding is also possible on the QS model. By setting a desired value and
clicking an arrow (up, down, right, left) the stage will move according to the
entered value. In addition to the machine coordinate system, the stage can be
moved using the workpiece coordinate system.The user can measure the
workpiece without worrying about its positioning on the table.
Stage movement in the workpiece
coordinate system
In addition to the machine coordinate system,
when moving the stage by step feeding, the
workpiece coordinate system is available.
The stage can be moved along the workpiece
coordinate system set.
Even though the workpiece is not placed in
parallel with the machine coordinate system,
after setting the workpiece coordinate system,
the stage can be moved along the workpiece
coordinate system.
Stage control through
graphic window
Stage control through
video window
Click the mouse with “Ctrl”
key, Quick Scope moves to the
position clicked.
Click the mouse with “Ctrl”
key, the clicked point will be
centered on the display.
Center coordinate
X: 0.000mm,
Y: 0.000mm
Center coordinate
X: 2.500mm,
Y: 0.000mm
Center coordinate
X: 5.000mm,
Y: 0.000mm
Repeat measurement navigation
(QS-E model)
The Quick Scope QS-L can repeat measurement routines (measuring commands,
illumination settings, zoom lens magnification, etc.) set in the Learn mode.
The distance from the current position (green cross-hairs at the center) to the
next target point (red cross-hairs) are indicated by the counter above the screen.
When moving to the next measurement point, all the operator has to do is move
the X,Y table until either the counter reads zero or by laying the green cross-hairs
on top of the red cross-hairs.
Start repeat-mode.
Move the stage to place the
green cross-hairs on top of
the red cross-hairs.
As the green cross-hairs
approach the red cross-hairs
the next measurement point
comes into the screen.
Current position
Target position
1
2
3
QSPAK Software